Original paperLocalization accuracy of two electromagnetic tracking systems in prostate cancer radiotherapy: A comparison with fiducial marker based kilovoltage imaging
Introduction
Accurate treatment localization is crucial part of modern radiotherapy. Its importance is emphasized in prostate radiotherapy as the prostate position is known to change relative to bony structures and skin surface between the fractions (interfraction motion) [1], [2], [3] and during the treatment (intrafraction motion) [4], [5], [6]. The motion is mainly caused by variations in rectum and bladder filling and it can range up to 1–2 cm during the treatment course [7], intrafraction motion being generally smaller in extent, typically less than 5 mm [5], [6]. Inaccuracies in the localization may deteriorate the target dose distribution [8]. Uncertainties in treatment localization, target delineation and treatment delivery are covered by planning target volume (PTV) margins around the clinical target volume (CTV). However, PTV margins expand the irradiated volume thus exposing larger volumes of nearby critical organs to high doses. High doses in rectum and bladder (especially trigone region) has been associated with late rectal bleeding [9] and incidence of acute and late urinary toxicity [9], [10], respectively. Uncertainties in localization, and thus PTV margins, could be reduced by accurate and frequently performed localization methods.
Several localization methods, such as ultra-sound (US), kilovoltage (kV) and megavoltage (MV) imaging, cone beam computed tomography (CBCT) and the use of fiducial markers (FM) have been applied in image-guided radiotherapy (IGRT) of the prostate. US methods increase the localization accuracy when compared to positioning based only on skin marks [11] but clinically unacceptable accuracy have also been reported [12]. US probe pressure can cause prostate displacement if handled without care, which have been seen both for transabdmominal [13] and transperineal probes [14]. Several studies have compared the accuracy of US localization methods to kV and MV imaging using FMs and found latter methods outperforming the US [11], [15], [16]. MV and kV imaging as such have poor soft tissue contrast and positioning of the patient is based on alignment of bony structures. CBCT imaging provides three-dimensional (3D) information of the imaging volume but the poor soft tissue contrast limits the accuracy of identifying the prostate from surrounding tissues. FMs implanted into the prostate provide a surrogate for prostate position, as they are made of radiopaque material and are discerned well in kV or MV images. Implantation of the FMs is made prior to treatment planning and the localization is based on alignment of the markers in reference images and kV or MV images (2D match) or markers in planning CT and CBCT images (3D match). FMs have been in use since the 1990s and the feasibility of FMs in the IGRT of the prostate has been shown in many studies [2], [17], [18]. Imaging of FMs has proven to be reliable localization method having accuracy on the order of ≤1 mm [19] which reduces the systematic positioning error [20] and enables the margin reduction [21], [22]. The imaging of FMs together with 3D soft tissue analysis is currently the most effective and widely available localization technique in the IGRT of the prostate [23] and is established practice in many radiotherapy clinics.
To monitor continuously the organ intrafraction motion a real-time electromagnetic (EM) localization system, Calypso (Varian Medical Systems), was introduced [24], [25]. Calypso consists of three transponders implanted into the prostate and an EM source/receiver array which is setup above the patient during the treatment. Calypso provides real-time 3D position information of the three transponders and thus the prostate and can be used for localization and intrafraction motion tracking. Another EM tracking system currently in the market is RayPilot (Micropos Medical AB), which consists of a wired transmitter, implanted into the prostate for the treatment course and removed afterwards, and a detector array which is setup on the treatment couch [26]. RayPilot provides the 3D position of the transmitter in real-time and can also be used in localization and tracking of the prostate. EM systems provide a non-ionizing alternative to kV- and MV-imaging systems and high sampling rate of the systems guarantees that positional information of the prostate is not missed during the treatment.
The localization accuracy of Calypso have been investigated in many studies both in laboratory conditions [19], [25], [27], [28], [29] and in clinical situations [4], [5], [30], [31], [32], [33] whereas there is a lack of information about the localization accuracy of RayPilot in clinical environment. Kindblom et al. [26] reported submillimeter accuracy of RayPilot in phantom fixture and compared 3D positional difference between RayPilot and orthogonal X-ray imaging when the transmitter was implanted in a urethral catheter. However, urethral catheter implantation does not represent the intended application method of implanting the transmitter into the prostate gland.
The aim of this study is to investigate the localization accuracy of the RayPilot and Calypso systems in comparison with fiducial marker based localization for a group of clinically treated prostate cancer patients. This is accomplished by comparing the translational couch shifts, proposed by the EM systems and couch shifts proposed by FM based kV-image alignment acquired simultaneously with the EM systems. Couch shifts represent the isocenter offsets recorded by the localization method in question. The stability and migration of the RayPilot transmitter, the gold seed fiducials and Calypso transponders are evaluated as well. To our knowledge, this is the first large study investigating the accuracy of the RayPilot system in clinical environment. The study is part of a clinical trial (ClinicalTrials.gov NCT02319239) which aims at developing extremely hypofractionated treatment protocol for prostate cancer.
Section snippets
RayPilot tracking system
The RayPilot system is described in more detail in a previous study [34]. The system consists of a wired transmitter implanted into the prostate, a receiver plate positioned on the treatment couch and software for the transmitter position evaluation. Transmitter signal is read by the receiver plate which is calibrated to the machine isocenter and based on the coordinates of the transmitter coil center point (CP) and treatment isocenter in the planning CT, the system can locate the correct
Difference in couch shifts
A total of 582 paired data points from 582 RayPilot fractions were included in the Bland-Altman analysis for each translational direction. The number of Calypso fractions analyzed was 335. For some of the fractions one or both of the kV projections had to be repeated and these extra data points with corresponding Calypso readings were included in the analysis as well. SI information could be read from both orthogonal kV projections, whereas AP and LR information could be acquired only from
Discussion
In the current study the positioning accuracy of RayPilot and Calypso positioning systems was evaluated against the orthogonal kV imaging using fiducial markers by comparing the couch shifts proposed by the different methods.
Bland-Altman analysis shows that the differences between the RayPilot and kV imaging were largely dispersed in AP (range −6.1, 8.6 mm) and SI (range −10.3, 7.0 mm) directions. In addition to this, a notable mean difference (−2.2 ± 2.4 mm) was seen in SI direction indicating
Conclusions
The results of the current study indicate that, mainly due to the instability of the transmitter, the localization accuracy of the RayPilot system in clinical circumstances is not equivalent to kV imaging of fiducial markers and is not adequate for interfraction localization of the prostate as such. RayPilot could be used for intrafractional motion tracking, but the initial localization should be made by some other means such as kV or CBCT imaging of fiducial markers. Instead, the results imply
Acknowledgements
This study was financially supported by the Seppo Nieminen funding, Grant number 15012, Tampere University Hospital. H Syrén is an employee of Micropos Medical AB.
References (47)
- et al.
Measurement of prostate movement over the course of routine radiotherapy using implanted markers
Int J Radiat Oncol Biol Phys
(1995) - et al.
Prostate motion during standard radiotherapy as assessed by fiducial markers
Radiother Oncol
(1995) - et al.
Evaluating the effect of rectal distension and rectal movement on prostate gland position using cine MRI
Int J Radiat Oncol Biol Phys
(1999) - et al.
Multi-institutional clinical experience with the Calypso system in localization and continuous, real-time monitoring of the prostate gland during external radiotherapy
Int J Radiat Oncol Biol Phys
(2007) - et al.
Observations on real-time prostate gland motion using electromagnetic tracking
Int J Radiat Oncol Biol Phys
(2008) - et al.
Organ motion and its management
Int J Radiat Oncol Biol Phys
(2001) - et al.
Image guidance in clinical practice – influence of positioning inaccuracy on the dose distribution for prostate cancer
Phys Med
(2018) - et al.
Predicting toxicity in radiotherapy for prostate cancer
Phys Med
(2016) - et al.
Bladder spatial-dose descriptors correlate with acute urinary toxicity after radiation therapy for prostate cancer
Phys Med
(2016) - et al.
Comparison of daily megavoltage electronic portal imaging with marker seeds to ultrasound imaging or skin marks for prostate localization and treatment positioning in patients with prostate cancer
Int J Radiat Oncol Biol Phys
(2006)
Comparison of ultrasound and implanted seed marker prostate localization methods: implications for image-guided radiotherapy
Int J Radiat Oncol Biol Phys
Technical aspects of daily online positioning of the prostate for three-dimensional conformal radiotherapy using and electronic portal imaging device
Int J Radiat Oncol Biol Phys
Analysis of fiducial marker-based position verification in the external beam radiotherapy of patients with prostate cancer
Radiother Oncol
Fiducial-based translational localization accuracy of electromagnetic tracking system and on-board kilovoltage imaging system
Int J Radiat Oncol Biol Phys
A comparison of the use of bony anatomy and internal markers for offline verification and an evaluation of the potential benefit of online and offline verification protocols for prostate radiotherapy
Int J Radiat Oncol Biol Phys
Prostate position relative to pelvic bony anatomy based on intraprostatic gold markers and electronic portal imaging
Int J Radiat Oncol Biol Phys
Intensity-modulated radiotherapy using implanted fiducial markers with daily portal imaging: assessment of prostate organ motion
Int J Radiat Oncol Biol Phys
A new system to perform continuous target tracking for radiation and surgery using non-ionizing alternating current electromagnetics
Int J Comput Assisted Radiol Surg
Accuracy of a wireless localization system for radiotherapy
Int J Radiat Oncol Biol Phys
High precision transponder localization using a novel electromagnetic positioning system in patients with localized prostate cancer
Radiother Oncol
The effect of transponder motion on the accuracy of the Calypso electromagnetic localization system
Int J Radiat Oncol Biol Phys
Target localization and real-time tracking using the Calypso 4D localization system in patients with localized prostate cancer
Int J Radiat Oncol Biol Phys
Positional stability of electromagnetic transponders used for prostate localization and continuous, real-time tracking
Int J Radiat Oncol Biol Phys
Cited by (12)
Monitoring Intrafraction Motion of the Prostate During Radiation Therapy: Suggested Practice Points From a Focused Review
2024, Practical Radiation OncologyMagnetic tracking and positioning in endoscopy
2021, Magnetic Materials and Technologies for Medical ApplicationsImage-Guided Radiotherapy for Pelvic Cancers: A Review of Current Evidence and Clinical Utilisation
2020, Clinical OncologyCitation Excerpt :Therefore, margins, technique and match strategy in these cases must be considered carefully to ensure the target(s) remain within their prescribed dose region and the OARs are not. Advanced techniques, including real-time tracking [65–67] and adaptive radiotherapy [68,69], are being explored in prostate radiotherapy; however, these approaches are not currently utilised in routine radiotherapy practice [21]. Radiotherapy for bladder cancer may be delivered as an adjuvant treatment in early-stage bladder cancers (e.g. after trans urethral resection of bladder tumour (TURBT)) or as the primary treatment in patients not suitable for surgery (either inoperable or wishing to avoid cystectomy).
Dosimetric effect of intrafraction motion and different localization strategies in prostate SBRT
2020, Physica MedicaCitation Excerpt :Image-guidance of conventionally fractionated prostate RT typically corrects for interfraction variations in the prostate position. Different imaging modalities applied in prostate RT have been discussed previously [5]. Most effective and widely available technique, and established practice in many clinics, is the daily kV or cone beam computed tomography (CBCT) imaging of implanted fiducial markers (FMs) [6].
Acute Side-effects of Different Radiotherapy Treatment Schedules in Early Prostate Cancer
2022, Anticancer Research