RT Journal Article SR Electronic T1 Evaluation of a Manual ELISA Kit for Determination of HER2/neu in Serum of Breast Cancer Patients JF Anticancer Research JO Anticancer Res FD International Institute of Anticancer Research SP 3067 OP 3073 VO 28 IS 5B A1 ANNE-SOPHIE GAUCHEZ A1 NADIA RAVANEL A1 DANIÈLE VILLEMAIN A1 FRANÇOIS-XAVIER BRAND A1 DOMINIQUE PASQUIER A1 RAOUL PAYAN A1 MIREILLE MOUSSEAU YR 2008 UL http://ar.iiarjournals.org/content/28/5B/3067.abstract AB Background: Our aim was to conduct an analytical validation in a routine laboratory setting of the cerb-B2/c-neu ELISA assay kit from Calbiochem® used to measure the extracellular domain (ECD) of HER2/neu in the serum of breast cancer patients. Materials and Methods: The evaluation was based on three different production lots used in a routine laboratory setting. The reference value was based on a population of 217 patients with breast cancer not overexpressing HER2. Results: The detection limit, below that given by the manufacturer, was 0.34 ng ml-1 and the quantification limit was 0.90 ng ml-1. Reproducibility and repeatability were at least 95%, precision coefficients of variation varied between 6 and 8.5% and trueness measured by dilution tests and the standard additions method varied between 97 and 107% . The threshold was estimated at 5 ng ml-1. Conclusion: This technique presents satisfactory levels of accuracy for routine laboratory use. Copyright© 2008 International Institute of Anticancer Research (Dr. John G. Delinassios), All rights reserved