Abstract
Aim: We developed a vaginal immobilization device for external radiotherapy in gynaecological malignancies and evaluated its bowel dose-reduction effect during carbon-ion radiotherapy (CIRT) and intensity-modulated radiation therapy (IMRT) in patients with cervical cancer. Patients and Methods: Computed tomographic images obtained with and without the device in seven patients with cervical cancer were assessed. Treatment plans for CIRT and IMRT were generated, and dose–volume parameters (V20, V25, V35, and D2cc) of the rectum, sigmoidal colon, and bladder were evaluated. Results: The mean±standard deviation of the rectal volume in CIRT for V35 with and without the device were 2.1±2.1 and 13.6±4.4 ml, respectively, and those in IMRT were 2.0±2.2 and 13.7±3.8 ml, respectively; these values were significantly lower in CIRT and IMRT using this device. Conclusion: Using our novel vaginal immobilization device, high rectal doses were largely reduced in CIRT and IMRT.
- Carbon ion radiotherapy
- intensity-modulated radiation therapy
- cervical cancer
- vaginal immobilization device
- rectal dose
High-precision external-beam therapy for uterine cervical cancer ideally requires sparing the dose to organs at risk such as the rectum, bladder, sigmoidal colon, and pelvic bones as much as possible (1-3). For instance, it is possible to reduce the rectal dose by spacing the rectum from the target and precisely immobilizing the vaginal position.
Particle therapies such as carbon-ion radiotherapy (CIRT) and proton-beam therapy have an excellent dose conformity compared with photon-beam therapy because they have a better Bragg peak and sharper penumbra (4, 5). Intensity-modulated radiation therapy (IMRT) also has an excellent dose density (6-8). Such therapies can minimize damage to normal tissues while effectively concentrating damage to the target tumour.
Hydrogel spacers (9, 10) and balloon catheters (11) have been used to reduce rectal and bladder doses of brachytherapy in patients with gynaecological cancer. Although these devices are effective, external radiotherapy should be used to confirm their positions at the time of irradiation and the device material must be as reproducible and as solid as possible because external radiotherapy requires high positional reproducibility and durability for a treatment period of several weeks.
We developed an intravaginal device for immobilizing the cervical position and spacing the rectum from the tumour in patients with gynaecological malignancies. The device has two functional requirements: The first is to confirm the device's position using internal markers on X-ray images before irradiation, and the second is to confirm the irradiated position and dose using a GAFchromic film after irradiation. In this study, we evaluated the effects of spacing the target from the rectum and assessed the reduction of bowel dose in treatment planning for CIRT and IMRT in patients with uterine cervical cancer.
Patients and Methods
Patient selection. This study was approved by the Institutional Review Board at Gunma University Hospital (registration: UMIN000013340) and informed consent was obtained from each patient. The study was performed in accordance with the ethical guidelines and regulations. Informed consent was obtained from each patient. The inclusion criteria were i) Histologically confirmed uterine cervical cancer, ii) no vaginal tumour invasion, iii) treatment by whole-pelvic irradiation with carbon ions, and iv) computed tomographic (CT) images were available both with and without insertion of the intravaginal device. We evaluated the data of seven consecutive patients who underwent CIRT for treatment of cervical cancer with passive irradiation from October 2013 to April 2015 at our facility. The patients' characteristics and the distances between the cervical tumour and rectum are shown in Table I. The distance between the cervical tumour and rectum was calculated as the average of the shortest distance between the cervical tumour and rectum in the axial plane where both the cervical tumour and rectum existed.
Device shape. The device [Ohno-Kubota-Tashiro (OKT) device, Japan patent P2016-32595A 2016] was made of high-density polyethylene (density of 1.16 g/cm3), and its stopping power ratio measured by 380 MeV/u of energy with a passive carbon-ion beam at our facility was 1.023. The device is available in four sizes: SS, S, L, and LL (length: 125.4, 125.4, 139.3, and 159.3 mm, respectively; vertical width: 28.9, 33.9, 37.6, and 37.6 mm, respectively; and horizontal width: 26.8, 31.6, 35.1, and 35.1 mm, respectively). Photographs of an L-size device are shown in Figure 1. All devices contain three tungsten spheres of 1.5 mm in diameter to confirm the device position on X-ray fluoroscopy for patient positioning, as shown in Figure 2. The tungsten markers are located in separate cross sections to confirm their three-dimensional positions using orthogonal X-ray fluoroscopy, and no large artefacts are produced during CT scans. The device can be split in the vertical direction as shown in Figure 1C, and GAFchromic film is inserted to confirm the irradiated dose and position. Furthermore, the device has a fixing port on the rear end to reduce misalignment for CT or irradiation as shown in Figure 1D.
In this study, we did not assess the reproducibility of the device position and irradiated dose by GAFchromic film because the primary study focus was the dose-reduction effect of the device.
CT imaging. Treatment-planning CT images were acquired with an X-ray CT device (Aquilion LB, Self-Propelled; Canon Medical Systems, Tochigi, Japan) approximately at the end of the expiration phase as monitored with a respiratory gating system (AZ-733; Anzai Medical, Tokyo, Japan). CT images with and without insertion of the device were consecutively acquired for all patients.
Treatment planning. Target delineation: The gross tumour volume was delineated on treatment-planning CT images with reference to T2-weighted magnetic resonance imaging and gynaecological examination by two radiological oncologists. The clinical target volume (CTV) included the cervical tumour, uterus, parametrium, the upper ≥2 cm of the vagina, and the pelvic lymph node regions. The planning target volume (PTV) was created from the CTV by adding 3-mm margins in all directions. When necessary, however, the common iliac lymph node regions in the PTV were cut so that they entered the irradiation field for CIRT (15×15 cm) because the dose-reduction effect on the bowel with and without the device was primarily studied. Outer organ contours were delineated for the bladder, rectum, and sigmoidal colon. The rectosigmoidal flexure was used as the border between the rectum and sigmoidal colon.
CIRT: Treatment planning for CIRT was calculated using an XiO-N system (Elekta, Stockholm, Sweden and Mitsubishi Electric, Tokyo, Japan) that employed a pencil-beam algorithm and incorporated a dose engine for CIRT (K2-Dose). The irradiation field was generated with a passive irradiation method (12). The relative biological effectiveness (RBE) was included in the absorbed dose using a spread-out Bragg peak concept (13), and the clinical dose was defined as Grays RBE.
Two beam fields (anterior–posterior and posterior–anterior) were used in six fractions per field, and the prescription dose was 3.0 Gy RBE ×12, i.e. 36 Gy RBE. The dose distribution was calculated to achieve minimum doses covering 95% (D95) of the PTV >95%.
IMRT: In order to evaluate the differences in the parameters of the dose–volume histogram (DVH) in CIRT and IMRT, treatment planning for IMRT was also calculated using an eclipse treatment system (Varian Medical Systems, Palo Alto, CA, USA).
Nine beam fields were used, and the prescription dose was 3.0 Gy ×12, i.e. 36 Gy. Each field angle was chosen, and the dose distribution was calculated to achieve D95% of PTV >95% and maximum dose of PTV <110% while reducing the rectal dose as much as possible.
Evaluation of treatment planning. In order to evaluate the effectiveness of the device, the percentage volume of the rectum, sigmoidal colon, and bladder receiving 20, 25, 30, or 35 Gy RBE/Gy (V20, V25, V30, or V35, respectively) and the percentage of minimum doses covering 2 cc of the rectum, sigmoidal colon, and bladder (D2cc) were calculated. Gy RBE was used as the dose unit for CIRT, and Gy was used as the dose unit for IMRT. All DVH parameters were calculated with and without the device and compared using a t-test with a 0.05 significance level.
Results
Examples of dose distributions for CIRT and IMRT with and without the device are shown in Figure 3. The doses to the rectum, sigmoidal colon, and bladder in CIRT and IMRT are shown in Table II. In CIRT, for the rectum, all parameters except V20 were significantly reduced by using the device. For the sigmoidal colon, V30 and V35 were reduced by using the device although there were no significant differences. For the bladder, all parameters except D2cc were significantly reduced by using the device. In IMRT, for the rectum, all DVH parameters were significantly reduced by using the device. For the sigmoidal colon, there were no significant differences. For the bladder, V20 was significantly reduced by using the device. Each DVH parameter, and the DVHs for the rectum, sigmoidal colon, and bladder are shown in Figure 4.
Discussion
We evaluated the rectal dose-reduction effect of our vaginal insertion device, which spaced the cervical tumour from the bowel (especially the rectum), in patients undergoing external-beam radiotherapy for cervical cancer. From the validation in seven patients (shown in Table I), we found that the device was able to create a space of 6.5 mm; the median distance between the cervical tumour and rectum in patients without the device was 2.6 mm, and that with the device was 9.1 mm.
In the treatment-planning study, we found that the rectal dose for both CIRT and IMRT was significantly reduced using the device. Damato et al. reported that D2cc for the rectum decreased by 22% using hydrogel injection in brachytherapy (10). Although in our study the reduction of D2cc for the rectum was low, D2cc decreased by 7.2% for CIRT and by 7.8% for IMRT, the reduction effect for a high dose volume was good; V35 was reduced by 84.6% for CIRT and by 85.4% for IMRT using our device.
In our comparison between CIRT and IMRT, the rectal volume reductions at high doses (V25, V30, and V35) were identical. Conversely, reduction of the medium dose was observed as shown by the significant difference in V20 for IMRT (Table II). For the sigmoidal colon in CIRT, V30 and V35 were lower with than without the device despite the fact that the volume was larger with than without the device. The sigmoidal dose might be reduced in such cases, although there were no significant differences in the present study. For the bladder, there were significant differences in almost all cases for CIRT and in one case for IMRT. However, the cause of these differences was assumed to be the large differences in the bladder volumes because the CT images with and without device insertion were consecutively acquired without breaks. Thus, a dose-reduction effect in the bladder by the device was not found.
CIRT for passive irradiation was validated in this study. A similar effect would be observed in layer-stacking irradiation or scanning irradiation with CIRT, and these techniques might reduce the bowel dose more than passive irradiation because they have better conformity (14, 15). Additionally, a similar effect would be observed in proton therapy because the proton beams also exhibit a Bragg peak and sharp dose distribution (16-18).
When the cervical tumour and rectum are in proximity, as shown in Figure 3, the rectum receives a high dose if the treatment-planning dose is calculated with a normal margin. In such cases, the high dose to the rectum can largely be reduced by modifying the plan to avoid the rectal side. However, it is necessary to consider whether and how much of the plan to modify because the target dose would also be reduced.
In this study, the significance of the device was verified in patients without vaginal infiltration. If the tumour has infiltrated the lower part of the vagina, the whole vaginal region should be included in the CTV, and the rectal dose would thus be unlikely to decrease. If the tumour has infiltrated the anterior or posterior vaginal wall, a further individualized sparing effect might occur. However, accurate diagnosis and positional precision would be necessary in such cases.
Because the device has three markers, its position in the patient can be confirmed from X-ray fluoroscopy. Additionally, a GAFchromic film can be placed inside the device. In vivo dosimetry during CIRT or IMRT using this film is expected to be developed.
This study has certain limitations. Further analysis is necessary to elucidate the exact effect of the device because this study involved only seven patients. Moreover, this study was performed only for treatment planning. The changes in the bowel position that would occur with inter- or intrafractional position changes of the device are unclear. Because the bowel dose has a risk of increasing above the treatment-planning dose, verification is necessary.
Conclusion
We developed an intravaginal device for immobilizing the cervical position and spacing the rectum from tumour in patients with gynaecological malignancies. In this study, we validated the effectiveness of rectal dose reduction in RT of the uterine cervix. In CIRT and IMRT, high rectal doses were largely reduced when using the device. In IMRT, medium rectal doses were reduced when using the device.
Acknowledgements
The Authors thank the staff at GHMC for their valuable insights. The Authors also thank Nobuyuki Kamei at the D.I.P. Co., Ltd., Gunma, Japan for providing technical support. Finally, the Authors thank Angela Morben, DVM, ELS, from Edanz Group (www.edanzediting.com/ac), for editing a draft of this article.
Footnotes
Authors' Contributions
YK and TO coordinated the entire study. Patient clinical data collection were performed by YK, MK, KM, and KT. Data analysis was performed by YK and MS. The article was prepared by YK. Corrections and improvements were suggested by TO, NO, SN, MT, and TN. All Authors read and approved the final article.
This article is freely accessible online.
Conflicts of Interest
The Authors declare that they have no competing interests in regard to this study.
Availability of Data and Materials
Data are available upon reasonable request and with permission of Gunma University Heavy Ion Medical Center.
Funding
This research did not receive any specific grant from funding agencies in the public, commercial, or not-for-profit sectors.
- Received February 6, 2019.
- Revision received February 27, 2019.
- Accepted March 7, 2019.
- Copyright© 2019, International Institute of Anticancer Research (Dr. George J. Delinasios), All rights reserved