Abstract
Background/Aim: Many patients with head-and-neck cancers receive radiotherapy. Treatment planning can be very complex in case of dental fillings or implants that cause metal artefacts. Verification of dose distributions may be performed using specific phantoms. This study aimed to develop a 3D-printed phantom that can be produced easily and cost-effectively. Patients and Methods: The phantom was designed to allow fast adaption to a patient's individual situation with a particular focus on metal artefacts due to dental fillings. Bone and soft-tissue shells were 3D-printed and filled with tissue-equivalent materials. Results: Attenuation properties of the tissue-equivalent structures in the phantom corresponded well to the structures of real human anatomy. In magnetic resonance (MR)-imaging, useful signals of the materials in the phantom were obtained. Conclusion: The phantom met the requirements including equivalence with human tissues and can be useful for highly individual treatment planning in precision-radiotherapy of head-and-neck cancers. It can be also used for scientific issues related to MR-imaging.
The number of patients with head-and-neck cancer is constantly increasing (1). Most patients with locally advanced tumors receive radiotherapy alone or following surgery. Since comparatively high radiation doses of 60-70 Gy are generally required, the treatment can be associated with significant acute and late toxicities (2-5). These toxicities include xerostomia, which can be burdensome for the patients and lead to late sequelae such as radiation caries (6). Xerostomia following radiotherapy can be reduced with the use of modern precision techniques such as intensity-modulated radiotherapy (IMRT) and volume-modulated arc therapy (VMAT) (6, 7). In order to provide the optimal radiation treatment for a head-and-neck cancer patient, the patient's individual situation including his anatomy must be appropriately considered. Treatment planning and delivery become more complex if a patient has dental fillings and implants or titanium implants as part of reconstructive surgery (8). Particularly, verification of the distribution of the radiation dose, which is an important procedure of quality assurance in radiation oncology, can become very difficult. Usually, the verification is performed at the linear accelerator (LINAC) without the patient. The calculated dose distribution of the treatment plan and the dose distribution measured at the LINAC using the attached imaging system are compared. In case of dental fillings or implants, this way of conventional verification of the dose distribution is less precise. Fillings and implants absorb radiation leading to deviations of the dose distribution, which cannot be appropriately considered during conventional verification. In these situations, verification may be performed with a specific phantom for dosimetric measurements, which considers the individual situation of a specific patient.
Phantoms are already commercially available for quality assurance purposes in radiation therapy. However, these phantoms cannot be adapted to a patient's individual situation. Moreover, they are comparatively expensive. Due to the fast development of rapid prototyping, it recently has become possible to produce phantoms that can take into account specific requirements for individual patients and can be produced at reasonable costs. The present study aimed to develop a new phantom specifically for head-and-neck cancer patients, which considers the impact of metal artifacts caused by dental fillings on the distribution of the radiation dose in these patients (9-11).
Patients and Methods
Requirements. The phantom had to be anthropomorphic. In addition, the use of various dosimetry procedures should be practicable. Moreover, it should be possible to adapt the phantom quickly and cost-effectively to various individual situations. To meet these requirements, the phantom should not be built as one piece but in layers. This means that all required changes will affect only one or very few layers but not the whole phantom. As a result, it would not be required to rebuild the whole phantom for each different situation. The layered structure of the phantom would also allow the placement of radiochromic films in the interlayer spaces. These films allow dose measurements and comparisons with planned dose distributions (12). Due to the layer arrangement, fixing options must be created for the stabilization of the phantom and its reproducible storage.
A clinically relevant aspect that needs to be addressed by the phantom is the influence of metal artifacts caused by dental implants and fillings on the real dose distribution in radiotherapy of head-and-neck cancer patients. Therefore, the teeth of the phantom should be exchangeable easily and fast. This would ensure that measurements can be carried out, with or without metallic fillings, without major modifications of the phantom. The phantom should be usable for various issues related to the dose distribution of external beam radiotherapy, brachytherapy and stereotactic radiotherapy (13-18). Moreover, it should be usable for magnetic resonance (MR)-imaging.
Testing of the material. The material for the construction of the phantom should meet two requirements, i.e. physical properties equivalent to human tissue considering attenuation coefficients and Hounsfield units (HU) and an easy and fast manufacturing process. In the preliminary stages of the construction process, different materials were tested with respect to processing and attenuation properties. For these tests, a computer tomograph (CT) (Siemens Somatom Balance, Siemens AG, Erlangen, Germany) was used. Data acquisition was performed with the following device settings: image matrix of 512×512, tube voltage of 130 kV, tube current of 34 mA, exposure time of 1500 msec, focus size of 0.95 mm, slice thickness of 3 mm and the reconstruction kernel B30s. As bone equivalent, conventional gypsum plaster (Modellgips für Bau und Hobby, decotric GmbH, Hann. Muenden, Germany) was tested. The gypsum plaster used was mixed with water (2:1). After 24 h, the gypsum plaster was completely cured and dried. The generated CT-images were evaluated using ImageJ (https://imagej.nih.gov/ij/). There was a pixel-by-pixel analysis within a region of interest (ROI) (26.52×12.88) and the calculation of a mean HU value. As tissue equivalent, three different materials were tested for their suitability: two-component silicone rubber (Sidopal Abformsilikon 8130-T, Fiberglas Discount GmbH, Hohenbrunn, Germany), two-component polyurethane casting resin (Polyurethan Giessharz Resinpal 1818, Fiberglas Discount GmbH, Hohenbrunn, Germany) and construction silicone (MEM Universal Silikon, Bauchemie GmbH, Leer, Germany).
Development of the phantom. The basis for the manufacturing of the phantom was a CT-dataset of the X-CAT software phantom (19). The dataset corresponded to a male adult with minor anatomical simplifications. The image pixels contain information about the attenuation values μ. For further processing Hounsfield units were required.
Therefore, the dataset was read into MATLAB (The MathWorks, Inc., Natick, MA, USA, Version: R2015a, ReadData3D: D. Kroon, University of Twente, 2010) and the
Hounsfield units were calculated pixel by pixel. The row of pixels separating the upper and lower rows of teeth was used as baseline to divide the entire dataset into layers. This resulted in eleven 2 cm-layers and one 3 cm-layer. For extracting individual structures from the CT images, a 3D modeling software was required, which enabled subsequent storage of the structures in a readable format for a 3D printer. The open source program 3D Slicer [Open Source Initiative, Palo Alto, USA, (20)] was used for segmentation of the surface and the bone shell of the corresponding layers with the following threshold values T:
and
The low HU-limit for the soft tissues resulted from the fact that only the surface should be segmented and not the complete soft tissue area.
Based on a surface triangulation, the different segmentations were stored in a stereo-lithography (stl) format. When creating stl-files from anatomical structures, free triangulations and incorrectly closed surface sections may occur due to the irregular surface structure, which require further processing. Using the open source software Blender (Stichting Blender Foundation, Amsterdam, the Netherlands), holes in the structure were closed and free triangulations eliminated. In addition, a simple smoothing filter was used. With this filter, the triangulations could be smoothened by flattening the angles between adjacent faces.
The printed bone shells (Figure 1) were filled with bone equivalent material and combined with the corresponding printed surface. Subsequently, the combined layer was filled with tissue-equivalent material. Air-filled anatomical structures such as the trachea and paranasal sinuses were simulated with silicone rubber. This could be removed without residue after curing of the tissue-equivalent material (Figure 2).
To study the influence of metal artifacts due to dental fillings, a CT-dataset of a conventional dental model was generated (Siemens Somatom Balance, matrix of 512×512, tube voltage of 130 kV, tube current of 90 mA, exposure time of 1500 msec, focus size of 0.6 mm, slice thickness of 1 mm and reconstruction kernel H80s). The dataset was processed as described above and printed twice. The dental model was filled with a combination of conventional plaster and pure calcium phosphate. One dataset served as reference. The other one was processed to fill drill holes with lead as an equivalent to amalgam fillings (Figure 3).
To be able to assess the phantom for further validation after completion of the manufacturing process, CT and MRI datasets were generated. A Siemens Biograph mCT40 was used for generation of the CT images with the following device settings: image matrix of 512×512, tube voltage of 120 kV, tube current of 372 mA, exposure time of 500 msec, focus size of 1.2 mm, slice thickness of 1 mm and the reconstruction kernel H60f. The used MRI was a Philips Ingenia 3.0 T (Philips GmbH, Germany) and one sequence was generated: T2W_TSE, dS Head coil, matrix of 576×576, slice thickness of 4 mm, spin echo, TR: 3000 msec, TE: 80 msec.
Three 3D-printed bone slices that corresponded to layers 3 to 5 of the phantom. After additive manufacturing the bone slices still contained supporting material (white), which was removed later.
Combination of the printed soft tissue and bone components: Filling process of layer 6 including nasal sinuses still filled with silicone rubber.
Overview of the materials tested with respect to the suitability for the construction of the phantom.
Results
Testing of the material. The results with respect to the testing of the materials are summarized in Table I. The silicone rubber was slightly viscous and therefore difficult to process. The mixing of the two components (rubber and crosslinking agent) resulted in numerous small air bubbles that could not be completely removed and resulted in an inhomogeneous mass. The advantage of the rubber was the processing time of only 20 min. The mean CT value was 283 HU (range=253-323±35 HU). For an appropriate reproduction of soft tissues, this HU value was too high [usually about 50 HU, (21)]. The consistence of the polyurethane casting resin was liquid. The mixing of the two components (casting resin and hardener) showed no formation of air bubbles. The mean CT value was 54 HU (range=46-82±18 HU), which was equivalent to the value of 50 HU for human soft tissue (21). At the specified mixing ratio, the gypsum plaster revealed a range of CT values between 909 and 1343 (±217) HU and a mean value of 1055 HU. The calculated HU values correspond to those of bones in the human body (21).
Development of the phantom. The new phantom representing a prototype met the pre-defined requirements with respect to tissue-equivalence, i.e. the equivalence of the materials used for the phantom with human tissues (Table I). Moreover, the additive manufacturing of the individual structures and layers corresponded well to the generated stl-files. The attenuation properties of the different structures in the phantom corresponded in a simplified form to those of the human anatomy. Examples of CT-images are given in Figure 4. Moreover, as anticipated, the results of the MR-imaging measurements revealed a usable signal in the area of the polyurethane casting resin and a very weak signal in the area of the 3D-printed material and the gypsum plaster for a T2-weighting (Figure 5). Thus, the phantom can be also used to contribute to scientific issues related to MR-imaging.
Slice of the upper jaw with metal implants (one molar and two premolars) corresponding to layer 8.
A few minor problems regarding the construction of the phantom were observed. Since the spaces inside the printed bone shell were sometimes too narrow or even closed, the gypsum plaster did not reach all cavities as required. This problem mainly occurred in the area of the facial skull and the temples and resulted in air bubbles in the bone area. For the vertebral bodies, a similar problem was observed. Since the proportion of the 3D printed part was too large, only a small area could be appropriately filled with gypsum plaster. This problem can be easily solved with a true-to-scale enlargement of the phantom.
Furthermore, the printed cylinders required for fixation of the individual layers for stabilization of the phantom broke under minor mechanical stress. This phenomenon was only observed when the base of the cylinder was parallel to the printing plate. For cylinders printed horizontally (base of the cylinder perpendicular to the printing plate) no breaks were observed under mechanical stress. However, fixation of the layers using cylinders turned out to be problematic in general. When taking into account the weight of each layer of the phantom, the adhesive area of the cylinders was comparatively small. As a consequence, the cylinders became loose during the use of the phantom. This problem can also be solved easily; printed cylinders must be replaced by commercially available plastic cylinders.
Selected CT layers of the phantom in the bone window showing uniform filling of the bone structure and filling with polyurethane castin resin.
Transverse layers from different sections of the phantom (spin echo, T2-weighted): The 3D-printing material and gypsum plaster show a very weak signal as anticipated. The signal of the polyurethane casting resin is sufficient for imaging.
Discussion
A lot of research is performed to improve the prognoses of patients with locally advanced head-and-neck cancers that are often poor (22-25). These patients are generally treated with surgery of the primary tumor and lymph nodes followed by radiotherapy plus/minus chemotherapy or with definitive radio-chemotherapy alone (2-6). These multimodal treatments are often associated with significant toxicities and functional restrictions (2-7, 26-28). The most common acute toxicities that are related to radiotherapy and radio-chemotherapy include dermatitis and oral mucositis (2-5). If interruptions of the radiotherapy course are required or the concurrent chemotherapy cannot be administered with the complete planned dose, the patients' prognoses become worse (29, 30). Xerostomia is a significant late morbidity that can be a consequence of surgical resection of one or both parotid glands and/or radio(chemo)therapy (2-7, 26-31). This type of late toxicity can significantly impair the patients' quality of life and lead to other late morbidities. Therefore, it is one important goal of radiotherapy of head-and-neck cancers to spare at least one parotid gland in order to reduce the risk of xerostomia. Sparing of the parotid gland(s) can be achieved with the use of modern precision techniques, namely IMRT and VMAT (6, 7). However, the treatment planning can be quite challenging if the patient has dental fillings or implants. In these patients, metal artefacts may not allow proper calculation and verification of the distribution of the radiation dose, and may even lead to undetected high radiation doses in the parotid glands resulting in xerostomia (8). Verification of the dose distribution may be more appropriately performed when using a phantom. Commercially available phantoms are comparatively expensive and generally not adaptable to the specific situation of an individual patient including dental fillings and implants.
Therefore, we have developed a new anthropomorphic 3D-printed phantom specifically designed for the treatment planning in radiotherapy of head-and-neck cancers. The idea of using this type of phantom has been recently applied to other specific situations including MR-imaging issues, validation of brain SPECT analyses and end-to-end tests in ion radiotherapy but not to radiotherapy of head-an-neck cancers (32-34). The phantom developed in the present study is more cost-effective than commercially available phantoms and can be produced in every radiation oncology department if a 3D-printer is available. Moreover, its construction in layers allows fast adaptions to new situations by making changes to only one or a few layers rather than re-building the whole phantom. The new phantom met the pre-defined requirements with respect to tissue-equivalence (equivalence between the materials of the phantom and corresponding human tissues). In addition, the additive manufacturing of the individual structures and layers corresponded well to the generated stl-files. Since usable signals of the materials of the phantom on MR-imaging were found, the new phantom can also be used for MR-imaging, which is another important field of research (32). Just a few minor problems regarding the construction of the phantom were observed regarding the appropriate filling of the phantom with the gypsum plaster and the printed cylinders used for connecting the layers and for stabilization of the phantom. However, one has to be aware that this new phantom is a prototype. Fortunately, these minor problems can be easily solved and will, therefore, not affect the value of this phantom for individual treatment planning in head-and-neck cancer patients. Further development of the phantom will include dental implants and metal implants from reconstructive surgery.
In conclusion, additive manufacturing (3D-printing) proved to be suitable for the construction of a new phantom developed for treatment planning in radiotherapy of head-and-neck cancers. The new phantom met the postulated requirements including the pre-defined tissue-equivalence. Particularly due to its construction in layers, it can be useful for individual treatment planning in precision-radiotherapy of head-and-neck cancers. Moreover, the phantom can be used for future scientific issues related to MR-imaging.
Footnotes
Authors' Contributions
M.G., M.S., C.Z., F.C., D.R. and T.B. participated in the design of the study. M.G., M.S. and S.B. provided the data. M.G., M.S., C.Z., F.C., D.R. and T.B. performed the analyses and the interpretation of the data. M.G., M.S., C.Z. and D.R. drafted the manuscript, which was reviewed and approved in its final form by all Authors.
This article is freely accessible online.
Conflicts of Interest
On behalf of all Authors, the corresponding Author states that there are no conflicts of interest related to this study.
- Received October 29, 2019.
- Revision received November 4, 2019.
- Accepted November 5, 2019.
- Copyright© 2019, International Institute of Anticancer Research (Dr. George J. Delinasios), All rights reserved